Determine the electrical resistance of a given copper wire using a potentiometer.
To determine the electrical resistance of a given copper wire using a potentiometer by comparing the potential drops across the unknown resistance and a standard (known) resistance when the same current flows through both.
A potentiometer is a device used to measure potential difference accurately. It consists of a long uniform wire through which a steady current is passed from an external battery (the primary circuit). Because the wire has uniform cross-sectional area and resistivity, its resistance is distributed uniformly along its length. By Ohm’s law, when a constant current I flows through this wire, the potential drop across any segment is directly proportional to the length of that segment.
The potential gradient is defined as the fall of potential per unit length of the potentiometer wire. If the total EMF of the driver cell produces a potential difference V across the full length L of the wire, then the potential gradient is:
Any potential difference to be measured is balanced against an appropriate length of the potentiometer wire, where the potential drop equals k × l. At the balance point, no current flows through the galvanometer, making this a null method of measurement.
In this experiment, the unknown resistance X (copper wire) and a known standard resistance R are connected in series in a secondary circuit so that the same current i passes through both. Using a two-way key, we alternately connect the potentiometer’s galvanometer arm across X and across R.
When the galvanometer arm is connected across the unknown resistance X and the jockey is moved along the wire until the galvanometer reads zero, let the balance length be l₁. Then:
Similarly, when the galvanometer arm is connected across the known resistance R and a new balance length l₂ is found:
Dividing the two equations eliminates both the current i and the potential gradient k:
This experiment is rooted in two foundational principles of circuit theory:
The diagram below shows the potentiometer arrangement for comparing potential drops across the unknown resistance X (copper wire) and the known resistance R.
Standard resistance used: R = 1.00 Ω
Least count of scale: 0.1 cm
| S.No. | Balance length for X, l₁ (cm) | Balance length for R, l₂ (cm) | X = R × l₁/l₂ (Ω) |
|---|---|---|---|
| 1 | 162.5 | 250.0 | 0.650 |
| 2 | 158.3 | 245.2 | 0.646 |
| 3 | 170.0 | 260.8 | 0.652 |
| 4 | 155.7 | 238.0 | 0.654 |
| 5 | 165.4 | 253.6 | 0.652 |
Given: R = 1.00 Ω, l₁ = 162.5 cm, l₂ = 250.0 cm
X = R × (l₁ / l₂)
X = 1.00 × (162.5 / 250.0)
X = 1.00 × 0.650
X = 0.650 Ω
Xmean = (0.650 + 0.646 + 0.652 + 0.654 + 0.652) / 5
Xmean = 3.254 / 5
Xmean = 0.651 Ω
Maximum deviation from mean = |0.651 − 0.646| = 0.005 Ω
A conservative uncertainty of ±0.03 Ω accounts for both random scatter and systematic effects (scale reading, contact resistance).
X = 0.65 ± 0.03 Ω
This value was obtained using a potentiometer with a 1.00 Ω standard resistance by the method of comparing potential drops. The null-method measurement ensures that the result is free from errors arising due to galvanometer resistance.
A potentiometer works on the principle that when a constant current flows through a wire of uniform cross-section and composition, the potential drop across any length of the wire is directly proportional to that length. Mathematically, V = kl, where k is the potential gradient (potential drop per unit length) and l is the length of the wire segment.
This proportionality holds because a uniform wire has uniform resistance per unit length. By Ohm’s law, V = IR, and since R itself is proportional to length (R = ρl/A for uniform ρ and A), the potential drop is proportional to length when the current is constant.
The potentiometer is classified as a null instrument because at the balance point, no current is drawn from the circuit under test. This gives it theoretically infinite input impedance at balance, making it more accurate than a voltmeter for precision measurements.
The potential gradient of a potentiometer wire is the fall of potential (voltage drop) per unit length of the wire. It is denoted by k and expressed in volts per metre (V/m) or volts per centimetre (V/cm).
If the total potential difference across the full length L of the potentiometer wire is V, then:
k = V / L
The potential gradient can be adjusted by changing the current through the wire (using the rheostat) or by changing the total length of wire in use. A smaller potential gradient makes the potentiometer more sensitive, because a given potential difference corresponds to a longer balance length, allowing more precise location of the null point.
In practice, the potential gradient can be determined by balancing a known EMF (such as a standard cell) against a measured length of the wire. If a standard cell of EMF Es balances at length ls, then k = Es / ls.
A potentiometer is preferred over a voltmeter for measuring electromotive force (EMF) for the following reasons:
The potentiometer essentially compares an unknown potential difference against a known, calibrated reference, making it a fundamental standard for voltage measurement.
Both the potentiometer and the Wheatstone bridge are null-method instruments that use a galvanometer to detect the balance condition (zero current through the galvanometer arm).
In a Wheatstone bridge, four resistances P, Q, R, and S are arranged in a diamond, and at balance: P/Q = R/S. The unknown resistance is found by comparison with known resistances.
In this potentiometer experiment, we effectively compare two potential drops using lengths of a uniform wire. At balance: X/R = l₁/l₂. The ratio of balance lengths plays the same role as the ratio of bridge arms.
The potentiometer can be thought of as a linear (stretched-out) form of the Wheatstone bridge, where the two arms formed by the potentiometer wire segments (of lengths l₁ and L − l₁, or l₂ and L − l₂) replace two of the four bridge arms. Both instruments exploit the proportionality of potential drops to resistances when the same current flows through them.
A practical difference is that the potentiometer offers continuously variable comparison (by sliding the jockey) rather than discrete steps (by switching resistance boxes), which can give finer resolution.
The rheostat in the primary circuit of the potentiometer serves two important purposes:
Additionally, the rheostat protects the potentiometer wire and the battery from excessive current, preventing overheating and ensuring stable operation throughout the experiment.